Hi everyone, I'm relatively new to freeze drying since it recently became a project in my lab job. I spent the year doing research and then was able to finally use a machine in Korea since I was sent there for training by my job. I used an Operon freeze dryer, and once I got back to my office in the US I got a Harvest Right. A lot of info for harvest right seems vague and very user experiment-based so please let me know any tips or comments!
In Korea, we froze the samples to -40C. It took ~1hr to get reach it and then we had it sit for an additional hour. Then it dried for 22 hours. This part happened overnight and the manual is Korean so unfortunately I don't have the most detailed parameters on that end. I'm trying to replicate this process in Harvest right so I freeze to -30F. I saw that the minimum target temp is -40F, but because there is not a complete vacuum seal on the door (door is still closed and latched and flush against O-ring) during freezing my machine was having trouble reaching this temperature and found -30F to be a reliable temperature that is as low as possible for me.
**I also want to make sure if this is something people experience as well? During freezing, I'm able to push door in with literally like a hair of room so when I first experimented going as low temp as possible, droplets were able to fall through O-ring down the door on the inside.
**2nd, my machine had no problem reaching -30F and 450mT for target temp and pressure in vacfreeze and whatever the rest of the default settings were for dry process and dry final. However, once i started manipulating target pressure, I haven't been able to go past vacfreeze at all. I tried 150mT target pressure, then 250mT, all in vacfreeze. i experienced a lot of temperature increasing and therefore pressure couldnt drop properly. Or I swear the vacuum wasn't removing pressure as efficiently for the first 10-15 minutes which I'm theorizing is why temperature is increasing naturally? Because sometimes I just nudge the door against O-ring and then suddenly, I see the vacuum pull and now the door is completely completely sealed. Could this be a vacuum leak? I don't see or hear any issues with O-ring, oil pump (I use premier oil), etc.
- to add to that, I get a short UnableAchVac error in my cycles and I'm wondering if there really is an issue with the machine? It's new, only received in September/Oct, only ran about 4 full cycles. I've had to stop a few mid cycle and just let things naturally come back to atmospheric pressure and room temp, part of me is a little worried that that might affect the machine but unsure yet.
**I'm really struggling to find info about this online lol, in the diagnostics screen when you click on the leaf logo it shows 3 tray temperatures. where are these sensors? I have 7 trays and each tray temp is different from each other with the biggest deviation being 1 tray always around ~7F lower than the others.
**Note, I have tried these parameters both for samples on the trays and running an empty tray trial. My freezing phase/cycle is always consistent time wise and cooling rate, but its once I started touching pressure things got a little more complicated.
I feel as if I'm going to ask more questions on here as I continue to mess around with things, but please let me know if anybody has any comments! Thanks guys