Spectron 7 (AC-APM7) read 60.8% RH vs 85.0% on an ISO 17025-calibrated meter. Leaf VPD 0.52 vs ~0.10 kPa. Data + certs inside.
I bought a Spectron 7 retail and checked it side by side against two thermohygrometers calibrated Oct 6, 2026 by an ISO/IEC 17025 accredited lab (certs 6040-17082981 and 6040-17082465).
Same location, same time:
- Spectron 7: 60.8% RH
- Reference meter 2B: 85.0% RH
- Gap: 24.2 points
Leaf VPD:
- Spectron 7: 0.52 kPa
- True leaf VPD: ~0.10 kPa
Why this matters: if you let the camera's readings override your Controller AI probes, the controller may add humidity to a room that's already wet, or trigger fail-safes it doesn't need. Until you've checked it against a calibrated meter, I'd keep your probes in charge.
This is one unit in my setup (8-layer indoor saffron stack, cool/humid conditions). Has anyone else checked theirs against a calibrated reference? I'd like to know if it's unit-specific or consistent.
Not sponsored, no affiliate links, no free units.