BBO SPDC alignment: Unexpectedly high singles counts with an off-center 405 nm pump beam, but no confirmed coincidences
Hi everyone!
I'm an undergraduate physics researcher working with a Thorlabs quantum optics kit to generate 810 nm photon pairs through spontaneous parametric down-conversion (SPDC) using a 405 nm pump laser and BBO crystal.
Today, I encountered something particularly confusing while troubleshooting our optical alignment.
When we follow the recommended alignment procedure and center the 405 nm pump beam, our detector counts typically remain in the hundreds of counts per second, and we haven't been able to establish a convincing coincidence signal.
However, after adjusting mirrors M1 and M2, the pump beam became visibly displaced from its intended path. Instead of passing through the center of the alignment aperture, it appeared to strike near the upper-left edge.
Despite this apparent misalignment, we began observing substantially higher singles counts.
Configuration 1: Beamsplitter installed
The following measurements were recorded simultaneously:
- Detector A: Approximately 40,000 counts/s.
- Detector B: Approximately 3,500 counts/s.
- Detector T: Remained in the hundreds of counts/s.
- Pump on/off test: Detector B showed sharp decreases in its count rate when we turned off the 405 nm pump.
Configuration 2: Beamsplitter removed
- Detector A: Increased to approximately 80,000 counts/s.
Removing the beamsplitter approximately doubled Detector A's count rate. However, this does not necessarily establish that the detected light originated from SPDC.
Coincidence measurements: Despite these elevated singles rates, our exploratory T-A/T-B delay scans have not produced a convincing, repeatable coincidence peak.
These count rates are approximate readings recorded during the experiment, as our screenshots do not display numerical y-axis values.
Previous troubleshooting attempts
We've spent considerable time attempting to locate and collect the expected 810 nm photon pairs, including:
- Centering the 405 nm pump beam using the alignment target and iris.
- Adjusting the BBO crystal's orientation using its mounting knobs.
- Aligning the crystal using the reflected alignment laser spot.
- Repositioning the collection optics and adjusting the detector collection geometry.
- Rotating and removing the 405 nm half-wave plate between M1 and M2.
- Following the troubleshooting procedures in the Thorlabs documentation and instructional videos.
Interestingly, some of our more carefully aligned configurations have produced substantially lower count rates than the configuration in which the pump beam is visibly displaced.
My main question is: Why would an apparently misaligned 405 nm pump beam produce significantly higher singles counts than our nominally aligned configuration, without corresponding evidence of photon-pair coincidences?
I'm considering several possible explanations, including scattered pump light, fluorescence, or changes in spatial collection geometry. I'm also unsure why Detector A registers substantially more counts than Detector T when the beamsplitter is installed.
I recognize that high singles rates alone do not demonstrate successful SPDC, particularly without coincidence measurements.
For anyone experienced with SPDC, nonlinear optics, or single-photon detection:
- Could the displaced pump beam be introducing additional scattered light or fluorescence into the collection path?
- Could our nominal pump alignment be geometrically correct but poorly optimized for phase matching or collection of the down-converted photons?
- What would explain the substantial difference between Detectors A and T with the beamsplitter installed?
- What diagnostic tests would you prioritize to distinguish genuine 810 nm SPDC photons from optical background?
I've attached photographs showing the unusual pump beam position, the BBO crystal and alignment hardware, the detector singles measurements, and our exploratory coincidence scans.
I'd really appreciate any suggestions or insights. I'm still learning experimental quantum optics, and this has been a particularly interesting troubleshooting challenge!





32
u/theQuick_BrownFox 4d ago
You are completely misaligned and picking fluorescence from some other optics. That’s why you are getting so many single counts. I had this issue as well. The best way to re-align is to the reverse light process in a fiber and start from there. It is harder to align than expected.
15
u/Independent_Vast9279 4d ago
Not sure about this kit, but BBO alignment is a lot more complicated that simply getting the beam through an aperture. Phase matching requires correct axial orientation, angle tuning for wavelength and compensating walk off.
If you know all this already, great. If any of this is news I’d suggest more reading before you mess with the optics.
The old saying goes: a month in the lab can save you an hour in the library. I’ve always found this safe advice.
Source: PhD in optical materials, 25 years hands on experience with leadership positions in companies you likely know and have bought hardware from.
2
u/Deep-Path-4883 4d ago edited 4d ago
The signal always depends on the crystal angle, the angle of incidence of the laser beam, and, if you are using a pulsed system, the pulse duration (this last factor does not apply to CW systems).
Furthermore, any optical component can cause absorption or scattering, and if pulse duration is crucial you introduce positive chirp. It is therefore to be expected that you will lose some signal when using a beamsplitter. Differences in A and T can be simply missalignment or differences in the properties in the beamsplitter (there are different types of splitters and not all split 50/50).
I would also recommend using a bandpass filter after the BBO crystal to effectively block the incoming beam. Otherwise, you will not be measuring the output alone. I suggest first aligning the beam so that it is perfectly straight, then rotating and tilting the crystal until you achieve maximum intensity at the correct spectrum.
2
u/einstein1351 4d ago
I've used this kit a bunch. Ill do a thorough read and reply tmm!
1
u/TortoiseBoy92 4d ago
Have you guys got access to any or all of: dichroics/filters to separate pump and SPDC; Spectrometer covering the signal region; Power meter?
Ideally you want to perform alignment with a simple way of determining your signal power, so usually with a filter and power meter (though your counts should do the job if you've filtered correctly and don't have a bunch of stray light bouncing around)
1
u/Fun-Ordinary-9751 4d ago
If the crystal axis wasn’t cut correctly this would make sense. It doesn’t help that when you effectively tweak the angle that the input and output spots are no longer centered on both faces.