Hi everyone!
I'm an undergraduate physics researcher working with a Thorlabs quantum optics kit to generate 810 nm photon pairs through spontaneous parametric down-conversion (SPDC), using a 405 nm pump laser and a BBO crystal. I've been troubleshooting our alignment for a while now, but today I ran into something that's genuinely confusing me.
Normally, when we follow the recommended alignment procedure and center the 405 nm pump beam through the alignment target, our detectors register only hundreds of counts per second. We also haven't been able to establish a convincing coincidence signal.
However, after adjusting mirrors M1 and M2, the pump beam became visibly displaced from its intended path. Instead of passing through the center of the alignment aperture, it was hitting near the upper-left edge.
What's strange is that our singles counts increased significantly despite the beam being visibly misaligned.
With the beamsplitter installed, we measured the following simultaneously:
- Detector A: ~40,000 counts/s
- Detector B: ~3,500 counts/s
- Detector T: Still in the hundreds of counts/s
After removing the beamsplitter, Detector A increased to approximately 80,000 counts/s.
We also tried turning the 405 nm pump off and on, and Detector B's count rate dropped sharply when the pump was turned off. So at least part of what we're detecting appears to be pump-dependent.
The problem is that, despite these higher singles rates, we still haven't observed a convincing, repeatable coincidence peak in our exploratory delay scans.
We've already tried adjusting the BBO crystal orientation, centering the pump with an alignment target and iris, repositioning the collection optics, and rotating or removing the 405 nm half-wave plate. Interestingly, some of our more carefully aligned configurations produced significantly lower count rates.
I'm struggling to understand why a configuration where the pump beam is clearly off-center would produce so many more counts than one where everything appears properly aligned.
Could this simply be scattered 405 nm light or fluorescence reaching the detectors? Or could the change in mirror angles somehow be improving collection of the down-converted photons despite the apparent misalignment?
I'm also unsure why Detector A receives substantially more counts than Detector T with the beamsplitter installed.
I know higher singles counts don't necessarily mean we're detecting genuine SPDC photons, especially without coincidences. I'm mainly trying to figure out which measurements or alignment adjustments would help us distinguish between these possibilities.
I've included photographs of the beam position, BBO crystal, and detector measurements in my original r/Optics post if anyone wants to see the setup:
https://www.reddit.com/r/Optics/s/AgBfTsjycv
I'd really appreciate any insight from people who've worked with SPDC or similar optical setups. Thanks!