r/massspectrometry • u/IonThePrize1 • 19h ago
Anyone know the actual electrical loads from your LCMS systems?
We have a 100A emergency backup panel in our facility running 4 LCMS instruments (2x Thermo Altis Plus, 1x Exploris 240, 1x Ascend) and are looking to add a QTOF and a GCMS. I can get all the "maximum" amp draws from the site prep guides, but I'm wondering if anyone has measured the actual amp draw from an instrument. For example, our Ascend needs ~25A from what I can gather, but the UPS only shows it drawing 8-10A while scanning with the LC running. I'm just trying to estimate our budget before going through the unnecessarily difficult process of getting a line monitor installed.