r/Optics • • 8h ago

BBO SPDC alignment: Unexpectedly high singles counts with an off-center 405 nm pump beam, but no confirmed coincidences

Thumbnail
gallery
13 Upvotes

Hi everyone!

I'm an undergraduate physics researcher working with a Thorlabs quantum optics kit to generate 810 nm photon pairs through spontaneous parametric down-conversion (SPDC) using a 405 nm pump laser and BBO crystal.

Today, I encountered something particularly confusing while troubleshooting our optical alignment.

When we follow the recommended alignment procedure and center the 405 nm pump beam, our detector counts typically remain in the hundreds of counts per second, and we haven't been able to establish a convincing coincidence signal.

However, after adjusting mirrors M1 and M2, the pump beam became visibly displaced from its intended path. Instead of passing through the center of the alignment aperture, it appeared to strike near the upper-left edge.

Despite this apparent misalignment, we began observing substantially higher singles counts.

Configuration 1: Beamsplitter installed

The following measurements were recorded simultaneously:

  • Detector A: Approximately 40,000 counts/s.
  • Detector B: Approximately 3,500 counts/s.
  • Detector T: Remained in the hundreds of counts/s.
  • Pump on/off test: Detector B showed sharp decreases in its count rate when we turned off the 405 nm pump.

Configuration 2: Beamsplitter removed

  • Detector A: Increased to approximately 80,000 counts/s.

Removing the beamsplitter approximately doubled Detector A's count rate. However, this does not necessarily establish that the detected light originated from SPDC.

Coincidence measurements: Despite these elevated singles rates, our exploratory T-A/T-B delay scans have not produced a convincing, repeatable coincidence peak.

These count rates are approximate readings recorded during the experiment, as our screenshots do not display numerical y-axis values.

Previous troubleshooting attempts

We've spent considerable time attempting to locate and collect the expected 810 nm photon pairs, including:

  • Centering the 405 nm pump beam using the alignment target and iris.
  • Adjusting the BBO crystal's orientation using its mounting knobs.
  • Aligning the crystal using the reflected alignment laser spot.
  • Repositioning the collection optics and adjusting the detector collection geometry.
  • Rotating and removing the 405 nm half-wave plate between M1 and M2.
  • Following the troubleshooting procedures in the Thorlabs documentation and instructional videos.

Interestingly, some of our more carefully aligned configurations have produced substantially lower count rates than the configuration in which the pump beam is visibly displaced.

My main question is: Why would an apparently misaligned 405 nm pump beam produce significantly higher singles counts than our nominally aligned configuration, without corresponding evidence of photon-pair coincidences?

I'm considering several possible explanations, including scattered pump light, fluorescence, or changes in spatial collection geometry. I'm also unsure why Detector A registers substantially more counts than Detector T when the beamsplitter is installed.

I recognize that high singles rates alone do not demonstrate successful SPDC, particularly without coincidence measurements.

For anyone experienced with SPDC, nonlinear optics, or single-photon detection:

  1. Could the displaced pump beam be introducing additional scattered light or fluorescence into the collection path?
  2. Could our nominal pump alignment be geometrically correct but poorly optimized for phase matching or collection of the down-converted photons?
  3. What would explain the substantial difference between Detectors A and T with the beamsplitter installed?
  4. What diagnostic tests would you prioritize to distinguish genuine 810 nm SPDC photons from optical background?

I've attached photographs showing the unusual pump beam position, the BBO crystal and alignment hardware, the detector singles measurements, and our exploratory coincidence scans.

I'd really appreciate any suggestions or insights. I'm still learning experimental quantum optics, and this has been a particularly interesting troubleshooting challenge!


r/Optics • • 10h ago

Optical Technician Job post Physics Bachelors looking for advice

3 Upvotes

decided to take a gap year inbetween my physics BA and further education(MA/Phd) I’m looking to stay in the optics industry so I was planning to work as an optical technician before starting an MA in Optical Engineering/optics ( gpa was no good for phds) looking for some advice and if this sounds like a good plan or if there might be better ones. I have a few years of instrumentation and optics research experience but my grades are just barely above a 3.0.